The incoming line reactor for medical CT machine is placed in front of the isolation transformer to improve the power factor;
Effectively limit the current surge caused by sudden change of grid voltage and operating voltage;
Smoothing the spikes contained in the supply voltage;
Effectively improve transformer power factor.
1、Low loss,low temperature rise,and the operation is safe and reliable;
2、Comply with relevant IEC6060-1-1 Medical safety standards;
3、Comply with relevant IEC6060-1-1 Medical EMC standards;
4、Excellent overload capacity,and low noise;
5、Suitable for wide range of voltage input,strong anti-saturation ability.
1、Altitude: no more than 2000m;
2、 Operating ambient temperature: -10℃~+40℃;
3、Relative humidity:<95%.
4、Maximum temperature gradient 6K/h;
5、Maximum absolute humidity 30 g/m3;
6、Atmospheric pressure range:800 to 1060? hPa。
1、 Rated voltage: 3Φ480V;
2、Rated frequency: 50Hz/60Hz;
3、 Rated current: 20A~500A;
4、Insulation class: Class H ;
5、Protection grade: IP00.